Acquisition Phase

Phase Included

Scan Delay

Respiration Phase

Anatomical Coverage

Non Contrast

N/A

N/A

N/A

N/A

Arterial Phase

Yes

Bolus triggered

Inspiration

Lung apices through the diaphragm to include the adrenal glands

Venous Phase

N/A

N/A

N/A

N/A

Delayed Phase

N/A

N/A

N/A

N/A

Scan Comments: A flash spiral mode can be used to improve scan speed and limit breathing artifact if the patient is short of breath. All scans are performed using ALARA principle.

 

Technical Parameters

Parameters

kVp

Care kV (ref. kV 120)

Effective mAs

CareDose

Care Dose Reference mAs

170

Time (Rotation)

0.28 sec

Average Acquisition Time

1-2 sec

Collimation

128 x 0.6mm

Pitch Value

2.2

Scan Direction

Craniocaudal

Comments: CareDose is used to reduce patient dose. 100 kVp can also be used for most average size patients which will significantly reduce patient dose.

 

Reconstruction Parameters

Soft Tissue

Thin Data

Lung

Slice Thickness

 3mm

 0.75mm

 3mm

Reconstruction Spacing

 3mm

 0.5mm

 3mm

Reconstruction Algorithm

 B30f

 B20f

 B80f

Window Width and Level

 550/50

 550/50

 1600/-500

Reconstruction Comments: When using Flash spiral mode, the DFOV is limited by the smaller of the two detector arrays to 33.2cm. This limitation can be overcome by reducing the pitch value to allow for a greater DFOV. Coronal and sagittal image reconstructions are also performed.

 

Contrast Parameters

Parameters

Contrast Type

Non ionic

Contrast Volume

80-100mL

Saline Flush

N/A

Injection Rate

5-6mL/sec

Oral Contrast

N/A

Contrast Volume

N/A

Comments: Bolus tracking is performed off the pulmonary trunk using a 200 HU trigger setting.