Acquisition Phase

Phase Included

Scan Delay

Respiration Phase

Anatomical Coverage

Non Contrast

Yes

N/A

Hold

Maxillary sinus through frontal sinus

Arterial Phase

N/A

N/A

N/A

N/A

Venous Phase

N/A

N/A

N/A

N/A

Delayed Phase

N/A

N/A

N/A

N/A

Scan Comments: IV contrast is not routinely necessary. Isocentering the patient within the gantry is key to image quality. All scans are performed using the ALARA principle.

 

Technical Parameters

Parameters

kVp

Care kV (ref. kV 100)

Effective mAs

CareDose

Care Dose Reference mAs

130

Time (Rotation)

1.0 sec

Average Acquisition Time

6-8 sec

Collimation

128 x 0.6mm

Pitch Value

0.8

Scan Direction

Caudocranial

Comments: CareDose is used to reduce patient dose.

 

Reconstruction Parameters

Soft Tissue

Thin Data

Bone

Slice Thickness

 3mm

 0.75mm

 3mm

Reconstruction Spacing

 3mm

 0.5mm

 3mm

Reconstruction Algorithm

 J30s

 J70h

 J70h

Window Width and Level

 350/20

 2500/500

 2500/500

Reconstruction Comments: Additionally, a soft tissue thin data set is performed using 0.75 mm x 0.5 mm slices, a J30s algorithm, and a window width and level of 350/20. Coronal and sagittal image reconstructions are also performed.

 

Contrast Parameters

Parameters

Contrast Type

N/A

Contrast Volume

N/A

Saline Flush

N/A

Injection Rate

N/A

Oral Contrast

N/A

Contrast Volume

N/A

Comments: IV contrast is not typically necessary for this exam.