Specific Anatomic RegionColon
ApplicationSuspected diverticulitis
AuthorFishman EK
Reference SourcePersonal Communication
Scanner UsedSiemens Plus4 Volume Zoom
KV / mAs / Time per Rotation (sec)120/165/.5
Detector Collimation (mm)2.5
Slice Thickness (mm)5
Data Reconstruction Interval (mm)5
Table Speed (mm per rotation) / Pitch12.5/5
Oral Contrast750-1000 cc of 3% Hypaque
IV Contrast Volume and Type120 cc of Omnipaque 350
Injection Rate2 cc/sec
Scan Delay (sec)50-60 sec
3D Technique UsedN/A
Comment:
Rectal contrast can be helpful, especially if a fistulae is suspected. We do not use rectal contrast routinely, although it could be done.