Siemens Definition 64 Slice MDCT Protocols: Head : Dental Adult

Source: Scanner default
Protocol Range Series Description kV Quality Ref. mAs (Eff.) mAs CAREDose/CAREDose 4D CAREDoseType CTDIvol RotTime Pitch Coll. Slice
Dental (Adult)Topogram Topogram 0.6 T20s 120 0 99 off CAREDose 0 0.500 1.0000000 0.60 0.60
 Dental  120 90 90 off CAREDose4D 12.105 1.000 0.90 0.60 
   Dental 0.75 H60s          0.75

 

Recon increment Images Kernel Window API Comment1
  1 T20s Topogram Head None 
     None 
 0.5 101 H60s Dental