Specific Anatomic RegionHead
ApplicationInner Ear
AuthorElliot K. Fishman, MD
Reference SourceJHU protocol
Scanner UsedSiemens Somatom Sensation 16
KV / Effective mAs / Rotation time (sec) 120 / 120 / 1.0
Detector Collimation (mm) 0.6
Slice Thickness (mm) 0.6
Feed / Scan (mm) 1.0
Kernel U90u
Increment (mm)4.0
Image Orderca-cr
Oral ContrastN/A
IV Contrast Volume and TypeN/A
Injection RateN/A
Scan Delay (sec)N/A
3D Technique UsedMPR is valuable with this isotropic data set.
Comment: -a key is to use the high resolution kernel for the reconstruction of data.