Specific anatomic region | Kidney |
Application | Renal donor |
Author | Elliot K Fishman, MD |
Reference Source | JHU Protocol |
Scanner Used | Siemens Sensation 64 |
KV/Effective mAs/ Rotation time (sec) | 120 / 250/ 0.33 |
Detector Collimation (mm) | 0.6 |
Slice thickness (mm) | 5.0mm |
Pitch | 0.75 |
Kernel | 30 |
Reconstruction interval | 5.0mm |
Image order (acquisition) | Cr-ca |
Oral contrast | 1000 cc water 15-20 minutes prior to CT |
IV contrast volume and type | 100-120 cc Omnipaque 350 |
Injection rate | 3-4cc/sec |
Scan delay (sec) | 25 sec (arterial)/ 60 sev (venous) |
3D technique used | VRT & MIP |
Comment: A 5 minute delayed topogram is obtained to get an "IVP" look at the renal pelvis and uterus. |