Purpose: To prospectively evaluate the effect of single- versus two segment image reconstruction on image quality and diag nostic accuracy at 64-section multidetector computed to mographic (CT) coronary angiography by using conven tional coronary angiography as the reference standard
Materials and Methods: The study design was approved by a human research com mittee; patients gave informed consent. The study was HIPAA compliant. Forty consecutive patients (22 men, 18 women; mean age, 61 years � 8 [standard deviation]) underwent both 64-section multidetector CT coronary angiography and conventional angiography. All data sets were reconstructed by using single- and two-segment image reconstruction algorithms, with resulting temporal resolution of 82.5-165 msec. Two experienced observers independently evaluated image quality and signs of coronary artery disease. A five-level grading scheme was used to grade stenosis (0%,
Results: Six hundred coronary artery segments were visible on conventional angiograms, of which 560 (93.3%) were seen by using single-segment and 561 (93.5%) were seen by using two-segment image reconstruction (P = .35). Mean quality scores were not significantly different (P ~ .22) for single- (3.1 � 0.9) and two-segment (3.2 � 0.8) reconstruction. Significantly (P = .03) better image quality was observed for two-segment reconstruction only at heart rates of 80-82 beats per minute, at which temporal resolution was approximately 83 msec. For grading coronary artery stenosis, correlation was 0.64 for single- and 0.66 for two-segment reconstruction (P = .43). Significant stenosis (>50%) was detected on a per-segment basis with 77.1% sensitivity and 98.6% specificity by using single-segment and with 79.2% sensitivity and 99.1% specificity by using two-segment image reconstruction.
Conclusion: At heart rates of more than 65 beats per minute, use of two-segment reconstruction improves image quality at multidetector CT coronary angiography but does not significantly affect overall diagnostic accuracy compared with single-segment