Acquisition Phase | Phase Included | Scan Delay | Respiration Phase | Anatomical Coverage |
Non Contrast | Yes | N/A | Inspiration | Limited to region of interest |
Arterial Phase | N/A | N/A | N/A | N/A |
Venous Phase | N/A | N/A | N/A | N/A |
Delayed Phase | N/A | N/A | N/A | N/A |
Scan Comments: IV contrast can be useful to evaluate swelling or abscess. All scans are performed using the ALARA principle. 3D post-processing is performed with this protocol. |
Technical Parameters | Parameters |
kVp | 100 / 140 |
Effective mAs | Use dose reduction software |
Quality Reference mAs | 250 / 193 |
Time (Rotation) | 0.5 sec |
Average Acquisition Time | 9-10 sec |
Collimation | 32 x 0.6mm |
Pitch Value | 0.6 |
Scan Direction | Craniocaudal |
Comments: Images are acquired using 100/140 kVp setting with a Sn filter. Dose reduction software is used to reduce patient dose. The pitch value should remain below 1.0 to reduce any artifacts during post-processing. |
Reconstruction Parameters | Soft Tissue | Bone | Thin Data | Thin Data |
Slice Thickness | 3 mm | 3 mm | 0.75 mm | 0.75 mm |
Reconstruction Spacing | 3 mm | 3 mm | 0.5 mm | 0.5 mm |
Reconstruction Algorithm | Q40f | I70f | Q30f | I70f |
Window Width and Level | 410/10 | 1776/176 | 410/10 | 1776/176 |
Reconstruction Comments: When generating mixed images, the DE composition is set to 0.5 and -1.0. Coronal and sagittal image reconstructions are also performed (3 mm x 3 mm, Q30f, 410/10 and 3 mm x 3 mm, I70f, 1776/176). |
Contrast Parameters | Parameters |
Contrast Type | Non-ionic |
Contrast Volume | 120 mL |
Saline Flush | N/A |
Injection Rate | 2-3 mL/sec |
Oral Contrast | N/A |
Contrast Volume | N/A |
Comments: IV contrast is not routinely necessary, however, if warranted we use the above protocol. |