EP No: 49
Scanogram - Lateral Scanogram Sternum to mid head | |
Scan mode | Scanogram(Lat) |
Start time(s) | 0 |
Wait time(s) | 0 |
kV | 120 |
mA | 150 |
Range(mm) | 500 |
Direction | OUT |
Display filter | Sharp (FL03) |
Scano angle ° | 90 |
CE | OFF |
Scanogram - AP Scanogram | |
Scan mode | Scanogram(AP) |
Start time(s) | 0 |
Wait time(s) | 6.1 |
kV | 120 |
mA | 50 |
Range(mm) | 500 |
Direction | IN |
Display filter | Sharp (FL03) |
Scano angle ° | 0 |
CE | OFF |
Helical - C3-T1 or as requested | |
Scan mode | Helical |
# of scan(s) | 1 |
Start time(s) | 0 |
Wait time(s) | 0 |
Collimation | 0.5 x 80 |
Pitch | Detail |
kV | 120 |
mA | Exp.3D Hi Quality |
Rotation time(s) | 0.5 |
Total scan time(s) | 4.6 |
Range(mm) | 180 |
Direction | IN |
CFOV | M |
CE | OFF |
CTDI(mGy) | *** |
DLP(mGy.cm) | *** |
Reconstruction | Axial1 | Axial2 | Volume1 | MutiView1 Coronal | MutiView1 Sagittal |
Slice thickness | 2.0 mm | 2.0 mm | 0.5 mm | 2.0 mm | 2.0 mm |
Slice Interval | 2.00 mm | 2.00 mm | 0.30 mm | 2.0 mm | 2.0 mm |
DFOV | 220 | 220 | 220 | *** | *** |
SUREIQ | Spine - Cervical | Spine - Bone | Spine - Cervical | Spine - Cervical | Spine - Cervical |
Recon Process | AIDR 3D Standard | AIDR 3D Standard | AIDR 3D Standard | *** | *** |
FC | 9 | 35 | 9 | *** | *** |
Boost 3D | OFF | OFF | OFF | *** | *** |
OSR | ON | ON | ON | *** | *** |
Filter | OFF | OFF | OFF | *** | *** |
* Exp.3D Adult Neck Hi Quality Preset - Reference Image Quality | |
Target SD | 7.5 |
Target Slice thickness | 3.0 mm |
Target SUREIQ | Neck Standard |
Recon Process | AIDR 3D Standard |
FC | 8 |
XY modulation | ON |
Max(mA) | 700 |
Min(mA) | 100 |