EP No: 55
Scanogram - Lateral Scanogram Sternum to mid head | |
Scan mode | Scanogram(Lat) |
Start time(s) | 0 |
Wait time(s) | 0 |
kV | 120 |
mA | 200 |
Range(mm) | 500 |
Direction | OUT |
Display filter | Sharp (FL03) |
Scano angle ° | 90 |
CE | OFF |
Scanogram - AP Scanogram | |
Scan mode | Scanogram(AP) |
Start time(s) | 0 |
Wait time(s) | 6.2 |
kV | 120 |
mA | 50 |
Range(mm) | 500 |
Direction | IN |
Display filter | Sharp (FL03) |
Scano angle ° | 0 |
CE | OFF |
S&V - Position in aortic arch Trigger at 120 HU | |
Scan mode | S&V |
# of scan(s) | 1 |
Start time(s) | *** |
Wait time(s) | 0 |
Collimation | 1.0 x 4 |
Pitch | *** |
kV | 100 |
mA | 50 |
Rotation time(s) | 0.275 |
Total scan time(s) | 0.275 |
Range(mm) | 4 |
Direction | OUT |
CFOV | M |
CE | OFF |
CTDI(mGy) | 1.5 |
DLP(mGy.cm) | 0.6 |
Reconstruction | Axial1 |
Slice thickness | 4.0 mm |
Slice Interval | *** |
DFOV | 260 |
SUREIQ | Neck - Low Dose |
Recon Process | OFF |
FC | 7 |
Boost 3D | OFF |
OSR | *** |
Filter | OFF |
* Exp.3D Adult Neck Quality Preset - Reference Image Quality | |
Target SD | |
Target Slice thickness | |
Target SUREIQ | |
Recon Process | |
FC | |
XY modulation | |
Max(mA) | |
Min(mA) |