EP No: 55

Scanogram - Lateral Scanogram Sternum to mid head
Scan modeScanogram(Lat)
Start time(s)0
Wait time(s)0
kV120
mA200
Range(mm)500
DirectionOUT
Display filterSharp (FL03)
Scano angle °90
CEOFF

 

Scanogram - AP Scanogram
Scan modeScanogram(AP)
Start time(s)0
Wait time(s)6.2
kV120
mA50
Range(mm)500
DirectionIN
Display filterSharp (FL03)
Scano angle °0
CEOFF

 

S&V - Position in aortic arch Trigger at 120 HU
Scan modeS&V
# of scan(s)1
Start time(s)***
Wait time(s)0
Collimation1.0 x 4
Pitch***
kV100
mA50
Rotation time(s)0.275
Total scan time(s)0.275
Range(mm)4
DirectionOUT
CFOVM
CEOFF
CTDI(mGy)1.5
DLP(mGy.cm)0.6
ReconstructionAxial1
Slice thickness4.0 mm
Slice Interval***
DFOV260
SUREIQNeck - Low Dose
Recon ProcessOFF
FC7
Boost 3DOFF
OSR***
FilterOFF

 

* Exp.3D Adult Neck Quality Preset - Reference Image Quality
Target SD 
Target Slice thickness 
Target SUREIQ 
Recon Process 
FC 
XY modulation 
Max(mA) 
Min(mA)