EP No: 54
Scanogram - Lateral Scanogram Sternum to mid head | |
Scan mode | Scanogram(Lat) |
Start time(s) | 0 |
Wait time(s) | 0 |
kV | 120 |
mA | 200 |
Range(mm) | 500 |
Direction | OUT |
Display filter | Sharp (FL03) |
Scano angle ° | 90 |
CE | OFF |
Scanogram - AP Scanogram | |
Scan mode | Scanogram(AP) |
Start time(s) | 0 |
Wait time(s) | 6.1 |
kV | 120 |
mA | 50 |
Range(mm) | 500 |
Direction | IN |
Display filter | Sharp (FL03) |
Scano angle ° | 0 |
CE | OFF |
S&V - Position below base of skull Manual trigger | |
Scan mode | S&V |
# of scan(s) | 1 |
Start time(s) | *** |
Wait time(s) | 0 |
Collimation | 1.0 x 4 |
Pitch | *** |
kV | 100 |
mA | 50 |
Rotation time(s) | 0.275 |
Total scan time(s) | 0.275 |
Range(mm) | 4 |
Direction | OUT |
CFOV | M |
CE | OFF |
CTDI(mGy) | 1.5 |
DLP(mGy.cm) | 0.6 |
Reconstruction | Axial1 |
Slice thickness | 4.0 mm |
Slice Interval | *** |
DFOV | 260 |
SUREIQ | Neck - Low Dose |
Recon Process | OFF |
FC | 7 |
Boost 3D | OFF |
OSR | *** |
Filter | OFF |
Helical - Mid head to below origins of arch vessels | |
Scan mode | Helical |
# of scan(s) | 1 |
Start time(s) | 0 |
Wait time(s) | 0 |
Collimation | 0.5 x 80 |
Pitch | Detail |
kV | 100 |
mA | Exp.3D Standard |
Rotation time(s) | 0.275 |
Total scan time(s) | 3.3 |
Range(mm) | 252 |
Direction | IN |
CFOV | M |
CE | ON |
CTDI(mGy) | *** |
DLP(mGy.cm) | *** |
Reconstruction | Axial1 | Volume1 |
Slice thickness | 3.0 mm | 0.5 mm |
Slice Interval | 3.00 mm | 0.30 mm |
DFOV | 260 | 260 |
SUREIQ | CTA - Neck | CTA - Neck |
Recon Process | AIDR 3D Standard | AIDR 3D Standard |
FC | 43 | 43 |
Boost 3D | OFF | OFF |
OSR | ON | ON |
Filter | OFF | OFF |
* Exp.3D Adult Neck Standard Preset - Reference Image Quality | |
Target SD | 12.5 |
Target Slice thickness | 3.0 mm |
Target SUREIQ | Neck Standard |
Recon Process | AIDR 3D Standard |
FC | 8 |
XY modulation | ON |
Max(mA) | 700 |
Min(mA) | 80 |