Harris GE Siemens
CTisUS Sponsers

Vascular CT: Trauma Abdomen CTA

SIEMENS FLASH DUAL SOURCE- UPDATED 1/1/12: Vascular

Acquisition Phase

Phase Included

Scan Delay

Respiration Phase

Anatomical Coverage

Non Contrast

 N/A

 N/A

N/A

N/A

Arterial Phase

 Yes

 bolus trigger

 inspiration

 Abdomen to include dome of liver through the pubic symphasis

Venous Phase

 N/A

 N/A

N/A

N/A

Delayed Phase

 N/A

 N/A

N/A

N/A

Scan Comments: In select cases a 60 -90 second delayed scan may be useful if aortic injury is detected on the arterial phase images. All scans are performed using the ALARA principle. All Scans are performed using the ALARA principle.

 

Technical Parameters

Parameters

kVp

120

Effective mAs

CareDose

Care Dose Reference mAs

290

Time (Rotation)

0.5 sec

Average Acquisition Time

5 -7sec

Collimation

128 x 0.6mm

Pitch Value

0.8

Scan Direction

craniocaudal

Comments: CareDose 4D is used to minimize patient exposure.

 

Reconstruction Parameters

Soft Tissue

Thin Data

Lung

Slice Thickness

 3mm

 0.75mm

 3mm

Reconstruction Spacing

 3mm

 0.5mm

 3mm

Reconstruction Algorithm

 B30F

 B20F

 B80F

Window Width and Level

 410/10

410/10

 1600/-500

Reconstruction Comments: Coronal and sagittal image reconstructions are also performed.

 

Contrast Parameters

Parameters

Contrast Type

 Non ionic

Contrast Volume

 120ml

Saline Flush

 N/A

Injection Rate

 4-5ml/sec

Oral Contrast

 H2O

Contrast Volume

 1000ml

Comments: Bolus Tracking is performed off the abdominal aorta using a 230 HU trigger setting.