Acquisition Phase | Phase Included | Scan Delay | Respiration Phase | Anatomical Coverage |
Non Contrast | Yes | N/A | Hold | Maxillary sinus through frontal sinus |
Arterial Phase | N/A | N/A | N/A | N/A |
Venous Phase | N/A | N/A | N/A | N/A |
Delayed Phase | N/A | N/A | N/A | N/A |
Scan Comments: IV contrast is not routinely necessary. All scans are performed using the ALARA principle. | ||||
Technical Parameters | Parameters |
kVp | Care kV (ref. kV 100) |
Effective mAs | CareDose |
Care Dose Reference mAs | 80 |
Time (Rotation) | 1.0 sec |
Average Acquisition Time | 3-5 sec |
Collimation | 128 x 0.6mm |
Pitch Value | 0.8 |
Scan Direction | Caudocranial |
Comments: Isocentering the patient within the gantry is key to image quality. 100 kVp is used to reduce patient exposure. CareDose is used to limit patient exposure. | |
Reconstruction Parameters | Soft Tissue | Thin Data | Bone |
Slice Thickness | 3mm | 0.75mm | 3mm |
Reconstruction Spacing | 3mm | 0.5mm | 3mm |
Reconstruction Algorithm | J30s | J70h | J70h |
Window Width and Level | 350/20 | 2500/500 | 2500/500 |
Reconstruction Comments: Additionally, a soft tissue thin data set is also performed using 0.75mm x 0.5mm slices with a J30s algorithm. Coronal and sagittal image reconstructions are also performed. | |||
Contrast Parameters | Parameters |
Contrast Type | N/A |
Contrast Volume | N/A |
Saline Flush | N/A |
Injection Rate | N/A |
Oral Contrast | N/A |
Contrast Volume | N/A |
Comments: IV contrast is not typically necessary for this exam. | |