Acquisition Phase | Phase Included | Scan Delay | Respiration Phase | Anatomical Coverage |
Non Contrast | N/A | N/A | N/A | N/A |
Arterial Phase | Yes | Bolus triggered | Inspiration | Aortic arch through Circle of Willis |
Venous Phase | N/A | N/A | N/A | N/A |
Delayed Phase | N/A | N/A | N/A | N/A |
Scan Comments: If bone removal is necessary, a Dual Energy CTA Carotid protocol should be used to optimize post-processing. All scans are performed using the ALARA principle. 3D post-processing is performed with this protocol. |
Technical Parameters | Parameters |
kVp | 120 |
Effective mAs | Use dose reduction software |
Quality Reference mAs | 300 |
Time (Rotation) | 0.33 sec |
Average Acquisition Time | 8-10 sec |
Collimation | 128 x 0.6mm |
Pitch Value | 0.35 |
Scan Direction | Caudocranial |
Comments: Dose reduction software is used to reduce patient dose and improve image quality, specifically in the shoulder region. |
Reconstruction Parameters | Soft Tissue | Bone | Hi Res Brain Only | Thin Data COW Mag |
Slice Thickness | 3 mm | 3 mm | 0.75 mm | 0.75 mm |
Reconstruction Spacing | 3 mm | 3 mm | 0.5 mm | 0.5 mm |
Reconstruction Algorithm | B30f | B60f | B50f | B20f |
Window Width and Level | 350/50 | 2500/500 | 700/80 | 350/50 |
Reconstruction Comments: Additional radial VRT images are also performed. |
Reconstruction Parameters | Thin Data Carotid Mag | Sliding MIP Axial | Sliding MIP Coronal | Sliding MIP Sagittal |
Slice Thickness | 0.75 mm | 16 mm | 16 mm | 16 mm |
Reconstruction Spacing | 0.5 mm | 3 mm | 3 mm | 3 mm |
Reconstruction Algorithm | B20f | H50f | H50f | H50f |
Window Width and Level | 350/50 | 700/80 | 700/80 | 700/80 |
Reconstruction Comments: Additional radial VRT images are also performed. |
Contrast Parameters | Parameters |
Contrast Type | Non-ionic |
Contrast Volume | 80-100 mL |
Saline Flush | N/A |
Injection Rate | 5-6 mL/sec |
Oral Contrast | N/A |
Contrast Volume | N/A |
Comments: Bolus tracking is performed off the aortic arch using a 25 HU trigger setting. |