Acquisition Phase | Phase Included | Scan Delay | Respiration Phase | Anatomical Coverage |
Non Contrast | N/A | N/A | N/A | N/A |
Arterial Phase | Yes | Bolus triggered | Inspiration | Aortic arch through Circle of Willis |
Venous Phase | N/A | N/A | N/A | N/A |
Delayed Phase | N/A | N/A | N/A | N/A |
Scan Comments: If bone removal is necessary, a Dual Energy CTA Carotid protocol should be used to optimize post-processing. All scans are performed using the ALARA principle. 3D post-processing is performed with this protocol. |
Technical Parameters | Parameters |
kVp | 100 / 140 |
Effective mAs | Use dose reduction software |
Quality Reference mAs | 270 / 270 |
Time (Rotation) | 0.28 sec |
Average Acquisition Time | 5-7 sec |
Collimation | 128 x 0.6mm |
Pitch Value | 0.7 |
Scan Direction | Caudocranial |
Comments: Images are acquired using 100/140 kVp setting with a Sn filter. Dose reduction software is used to reduce patient dose. A 0.28 rotation time is used to increase scan speed to avoid venous contamination. |
Reconstruction Parameters | Soft Tissue Open Field | Bone Open Field | Thin Data Open Field | Hi Res Brain Only |
Slice Thickness | 3 mm | 3 mm | 0.75 mm | 0.75 mm |
Reconstruction Spacing | 3 mm | 3 mm | 0.5 mm | 0.5 mm |
Reconstruction Algorithm | D30f | B70f | Q30f | B50f |
Window Width and Level | 350/50 | 2500/500 | 350/50 | 700/80 |
Reconstruction Comments: When generating mixed images, the DE composition is set to 0.8. |
Reconstruction Parameters | Thin Data Carotid Mag | Sliding MIP True Axial (Brain Only) | Sliding MIP Coronal (Brain Only Aligned to Basilar Artery) | Sliding MIP Sagittal (Brain Only) |
Slice Thickness | 0.75 mm | 16 mm | 16 mm | 16 mm |
Reconstruction Spacing | 0.5 mm | 3 mm | 3 mm | 3 mm |
Reconstruction Algorithm | I26f | B50f | B50f | B50f |
Window Width and Level | 350/50 | 700/80 | 700/80 | 700/80 |
Reconstruction Comments: When generating mixed images, the DE composition is set to 0.8. |
Contrast Parameters | Parameters |
Contrast Type | Non-ionic |
Contrast Volume | 60-100 mL |
Saline Flush | N/A |
Injection Rate | 5-6 mL/sec |
Oral Contrast | N/A |
Contrast Volume | N/A |
Comments: Bolus tracking is performed off the aortic arch using a 25 HU trigger setting. |