Acquisition Phase | Phase Included | Scan Delay | Respiration Phase | Anatomical Coverage |
Non Contrast | N/A | N/A | N/A | N/A |
Arterial Phase | Yes | Bolus triggered | Inspiration | Entire lower extremity vasculature of affected area |
Venous Phase | N/A | N/A | N/A | N/A |
Delayed Phase | N/A | N/A | N/A | N/A |
Scan Comments: All scans are performed using the ALARA principle. 3D post-processing is performed with this protocol. | ||||
Technical Parameters | Parameters |
kVp | 120 |
Effective mAs | Use dose reduction software |
Quality Reference mAs | 250 |
Time (Rotation) | 0.5 sec |
Average Acquisition Time | 16-18 sec |
Collimation | 128 x 0.6mm |
Pitch Value | 0.8 |
Scan Direction | Craniocaudal |
Comments: Dose reduction software is used to reduce patient dose. | |
Reconstruction Parameters | Soft Tissue | Bone | Thin Data | Coronal |
Slice Thickness | 3 mm | 5 mm | 0.75 mm | 3 mm |
Reconstruction Spacing | 3 mm | 5 mm | 0.5 mm | 3 mm |
Reconstruction Algorithm | B30f | B70f | B20f | B30f |
Window Width and Level | 410/10 | 1776/176 | 410/10 | 410/10 |
Reconstruction Comments: Thin slice bone reconstructions (B70f) are performed when fracture is suspected. 3D VRT and MIP images are also performed. | ||||
Reconstruction Parameters | Sagittal | ||
Slice Thickness | 3 mm | ||
Reconstruction Spacing | 3 mm | ||
Reconstruction Algorithm | B30f | ||
Window Width and Level | 410/10 | ||
Reconstruction Comments: Thin slice bone reconstructions (B70f) are performed when fracture is suspected. 3D VRT and MIP images are also performed. | |||
Contrast Parameters | Parameters |
Contrast Type | Non-ionic |
Contrast Volume | 120 mL |
Saline Flush | N/A |
Injection Rate | 4-5 mL/sec |
Oral Contrast | N/A |
Contrast Volume | N/A |
Comments: Bolus tracking is performed off the abdominal aorta using a 230 HU trigger setting. | |