Acquisition Phase | Phase Included | Scan Delay | Respiration Phase | Anatomical Coverage |
Non Contrast | N/A | N/A | N/A | N/A |
Arterial Phase | N/A | N/A | N/A | N/A |
Venous Phase | Yes | 50 sec | N/A | Base of skull through vertex |
Delayed Phase | N/A | N/A | N/A | N/A |
Scan Comments: Non-contrast images are performed when a bleed is suspected prior to CTV. Using a dual energy protocol is useful if bone removal is warranted. Iso-centering the patient within the gantry is key to image quality. All scans are performed using the ALARA principle. | ||||
Technical Parameters | Parameters |
kVp | 100 / 140 |
Effective mAs | Use dose reduction software |
Care Dose Reference mAs | 270 / 270 |
Time (Rotation) | 0.33 sec |
Average Acquisition Time | 4-5 sec |
Collimation | 128 x 0.6mm |
Pitch Value | 0.55 |
Scan Direction | Caudocranial |
Comments: Iso-centering the patient within the gantry is key to image quality. Dose reduction software is used to reduce patient dose. | |
Reconstruction Parameters | Soft Tissue | Bone | Thin Data | Hi Res |
Slice Thickness | 3 mm | 3 mm | 0.75 mm | 0.75 mm |
Reconstruction Spacing | 3 mm | 3 mm | 0.5 mm | 0.5 mm |
Reconstruction Algorithm | D30f | B70f | I26f | B50f |
Window Width and Level | 350/50 | 2500/500 | 700/80 | 700/80 |
Reconstruction Comments: When generating mixed images, the DE composition is set to 0.8. | ||||
Reconstruction Parameters | Sliding MIP Axial | Sliding MIP Coronal | Sliding MIP Sagittal |
Slice Thickness | 16 mm | 16 mm | 16 mm |
Reconstruction Spacing | 3 mm | 3 mm | 3 mm |
Reconstruction Algorithm | B50f | B50f | B50f |
Window Width and Level | 700/80 | 700/80 | 700/80 |
Reconstruction Comments: When generating mixed images, the DE composition is set to 0.8. | |||
Contrast Parameters | Parameters |
Contrast Type | Non-ionic |
Contrast Volume | 120 mL |
Saline Flush | N/A |
Injection Rate | 4-5 mL/sec |
Oral Contrast | N/A |
Contrast Volume | N/A |
Comments: A fixed delay of 50 seconds is optimal. | |