Acquisition Phase | Phase Included | Scan Delay | Respiration Phase | Anatomical Coverage |
Non Contrast | N/A | N/A | N/A | N/A |
Arterial Phase | Yes | Bolus triggered | N/A | Base of skull through vertex |
Venous Phase | N/A | N/A | N/A | N/A |
Delayed Phase | N/A | N/A | N/A | N/A |
Scan Comments: Non-contrast images are performed when indicated prior to CTA. Using a dual energy protocol is useful if bone removal is warranted. We typically perform these studies using single energy because the acquisition speed is slightly faster allowing for less likelihood of venous contamination to occur. All scans are performed using the ALARA principle. 3D post-processing is performed with this protocol. | ||||
Technical Parameters | Parameters |
kVp | 120 |
Effective mAs | Use dose reduction software |
Care Dose Reference mAs | 300 |
Time (Rotation) | 0.5 sec |
Average Acquisition Time | 2-3 sec |
Collimation | 128 x 0.6mm |
Pitch Value | 0.8 |
Scan Direction | Caudocranial |
Comments: Iso-centering the patient within the gantry is key to image quality. Dose reduction software is used to reduce patient dose. | |
Reconstruction Parameters | Soft Tissue | Bone | Hi Res | Thin Data COW Mag |
Slice Thickness | 3 mm | 3 mm | 0.75 mm | 0.75 mm |
Reconstruction Spacing | 3 mm | 3 mm | 0.5 mm | 0.5 mm |
Reconstruction Algorithm | H30f | H60f | H50f | H20f |
Window Width and Level | 350/50 | 2500/500 | 700/80 | 350/50 |
Reconstruction Comments: Additional radial VRT images are also performed. | ||||
Reconstruction Parameters | Sliding MIP True Axial | Sliding MIP Coronal | Sliding MIP Sagittal |
Slice Thickness | 16 mm | 16 mm | 16 mm |
Reconstruction Spacing | 3 mm | 3 mm | 3 mm |
Reconstruction Algorithm | B50f | B50f | B50f |
Window Width and Level | 700/80 | 700/80 | 700/80 |
Reconstruction Comments: Additional radial VRT images are also performed. | |||
Contrast Parameters | Parameters |
Contrast Type | Non-ionic |
Contrast Volume | 50 mL |
Saline Flush | N/A |
Injection Rate | 5-6 mL/sec |
Oral Contrast | N/A |
Contrast Volume | N/A |
Comments: Bolus tracking is performed off the carotid arteries. | |