Acquisition Phase | Phase Included | Scan Delay | Respiration Phase | Anatomical Coverage |
Non Contrast | Yes | N/A | N/A | Bass of skull through vertex |
Arterial Phase | N/A | N/A | N/A | N/A |
Venous Phase | N/A | N/A | N/A | N/A |
Delayed Phase | N/A | N/A | N/A | N/A |
Scan Comments: The F/U shunt protocol is performed using a low dose sequential scan to limit patient exposure. Isocentering and properly positioning the patient within the gantry is key to image quality and limiting the radiation dose. If the patient is unable to hold still for the duration of the exam, a low dose spiral study can be performed instead to decrease the scan time. All scans are performed using the ALARA principle. | ||||
Technical Parameters | Parameters |
kVp | 120 |
Effective mAs | Use dose reduction software |
Care Dose Reference mAs | 150 |
Time (Rotation) | 1.0 sec (Full) / 3.0 sec cycle time |
Average Acquisition Time | 10 sec |
Collimation | 128 x 0.6mm |
Table Feed | 34.5 mm |
Scan Direction | Caudocranial |
Comments: Dose reduction software is used to reduce patient dose. The F/U shunt protocol is performed using a low dose sequential scan to limit patient exposure. | |
Reconstruction Parameters | Soft Tissue | Bone | Thin Data | |
Slice Thickness | 5 mm | 5 mm | 0.75 mm | |
Reconstruction Spacing | 5 mm | 5 mm | 0.5 mm | |
Reconstruction Algorithm | J30s (3) | J70h (3) | J30s (3) | |
Window Width and Level | 100/30 | 2500/500 | 100/30 | |
Reconstruction Comments: 3D MPR’s can be performed to correct positioning. | ||||
Contrast Parameters | Parameters |
Contrast Type | N/A |
Contrast Volume | N/A |
Saline Flush | N/A |
Injection Rate | N/A |
Oral Contrast | N/A |
Contrast Volume | N/A |
Comments: IV contrast is not typically necessary for this exam. | |