Specific Anatomic Region | Renal arteries |
Application | Renal artery stenosis evaluation |
Author | Smith, P.A. and Fishman E.K. |
Reference Source | Three-dimensional CT angiography: Renal applications. Semin US CT MR, 19(5), 413-424. |
Scanner Used | Siemens Somatom Plus or Plus-4 |
Injection Rate | 3ml/sec |
Contrast Volume and Type | 120-150cc Omnipaque 350 |
Area Scanned | Entire length of both kidneys |
Scan Delay | 25sec |
Length of Spiral (time) | 30-40sec |
Slice Thickness | 2mm |
Table Speed/Pitch | 3-4mm/sec |
Reconstruction Interval | 1mm |
3D Technique Used | Volume rendering |
Comment: This protocol covers both kidneys and so can evaluate multiple renal aa. A set of noncontrast scans are obtained to first define the area of interest. |