Specific Anatomic Region | Spleen | |
Application | R/O splenic infarct | |
Author | Urban BA et al. | |
Reference Source | Helical CT of the spleen. AJR Am J Roentgenol. 1998 Apr; 170(4): 997-1003. | |
Scanner Used | Siemens Somatom Plus 4 | |
Injection Rate | 2cc/sec | |
Contrast Volume and Type | 100cc Omnipaque 350 | |
Area Scanned | Entire span of spleen | |
Scan Delay | 50sec | |
Length of Spiral (time) | 20-30sec | |
Slice Thickness | 5mm | |
Table Speed/Pitch | 5mm/sec | |
Reconstruction Interval | 5mm | |
3D Technique Used | None | |
Comment: Due to the dual blood supply of the spleen scans are not obtained early so that a 'pseudolesion' will not occur. |