| Specific Anatomic Region | Orbit | |
| Application | Trauma | |
| Author | Rhea JT et al. | |
| Reference Source | Radiol Clin NA Vol. 37 No. 3 May 1999: 489-513 | |
| Scanner Used | GE Hi-Lite Advantage | |
| Injection Rate | N/A | |
| Contrast Volume and Type | None used | |
| Area Scanned | Coronal- from front of eyelids to behind the optic canals. Axial- through orbits. | |
| Scan Delay | None | |
| Length of Spiral (time) | Not defined | |
| Slice Thickness | Coronal: 1mm Axial: 1mm | |
| Table Speed/Pitch | Coronal: 1.5mm/sec Axial: 1mm | |
| Reconstruction Interval | Coronal: 1mm Axial: 1mm | |
| 3D Technique Used | Shaded surface rendering | |
| Comments: 1.This protocol is also used for foreign body localization. 2. Images are filmed at both soft tissue and bone windows. | ||