| Specific Anatomic Region | Shoulder |
| Application | Trauma to shoulder joint (includes scapula and humerus) |
| Author | Fishman EK |
| Reference Source | Personal communication |
| Scanner Used | Siemens Somatom Plus 4 |
| Injection Rate | N/A |
| Contrast Volume and Type | N/A |
| Area Scanned | From just above acromioclavicular joint through scapular tip |
| Scan Delay | None |
| Length of Spiral (time) | 40sec |
| Slice Thickness | 3mm |
| Table Speed/Pitch | 3-5mm/sec |
| Reconstruction Interval | 2mm |
| 3D Technique Used | Volume rendering and multiplanar obliques |
| Comment: The technique can be modified if only the humeral head is injured. In cases with complex trauma IV contrast may be used to exclude a vascular injury to the mediastinum. The protocol used would use 120ml of contrast injected at 2-3ml/sec with a 30sec delay. |