Specific Anatomic Region | Liver |
Application | Suspected hepatoma |
Author | Baron RL et al. |
Reference Source | Radiology 1996; 199:505-511 |
Scanner Used | GE HiSpeed Advantage |
Injection Rate | 3-5ml/sec |
Contrast Volume and Type | 150ml of Optiray 320 |
Area Scanned | Entire liver |
Scan Delay | Arterial phase (AP): 20secDelayed phase (DP): 60-70sec |
Length of Spiral (time) | 20-30sec each |
Slice Thickness | 7mm |
Table Speed/Pitch | 7mm/sec |
Reconstruction Interval | 7mm |
3D Technique Used | None |
Comment: If the scanner can only obtain a single spiral scan then the arterial-phase study should be chosen as this is the best phase for lesion detection. Some sites will also do non contrast studies although we typically do not find it necessary. |