| Specific Anatomic Region | Liver |
| Application | Suspected metastases from colon cancer |
| Author | Ch'en IY et al. |
| Reference Source | J Comput Assist Tomogr 21(3):391-397 |
| Scanner Used | Somatom Plus-4 or GE HiSpeed Advantage |
| Injection Rate | 5ml/sec |
| Contrast Volume and Type | 200ml of Omnipaque-300 |
| Area Scanned | Entire liver |
| Scan Delay | 25-30sec (arterial)60-70sec (portal venous) |
| Length of Spiral (time) | Up to 30sec |
| Slice Thickness | 5mm |
| Table Speed/Pitch | 5-8mm/sec or a pitch of 1-1.6 |
| Reconstruction Interval | 2.5mm |
| 3D Technique Used | None |
| Comment: Although a dual-phase study might in theory be better, our experience and that of Ch'en et al. shows little difference in actual results. Dual-phase spiral CT does not appear to be indicated in the evaluation of liver metastases from colon cancer. Portal-venous-phase imaging alone is necessary. |