Specific Anatomic Region | Kidney |
Application | Evaluation of living renal donor |
Author | Cochran ST, et al. |
Reference Source | AJR 1997; 168:1569-1573 |
Scanner Used | Siemens Somatom Plus S |
Injection Rate | 3.5ml/sec (mean) |
Contrast Volume and Type | 120-150ml (type not defined) |
Area Scanned | Entire kidneys |
Scan Delay | 20sec |
Length of Spiral (time) | Approx 32sec |
Slice Thickness | 3mm |
Table Speed/Pitch | 5 or 6mm/sec or pitch 1.7-2 |
Reconstruction Interval | 2mm |
3D Technique Used | MIP |
Comment: 10 minutes after contrast injection, conventional radiographs (AP, obliques) were obtained. A delayed topogram (scout view) on the CT scanner is usually enough. |