Specific Anatomic Region | Muscle | |
Application | R/O tumor | |
Author | Fishman EK | |
Reference Source | Personal Communication | |
Scanner Used | Siemens Plus4 Volume Zoom | |
KV / mAs / Time per Rotation (sec) | 120/165/.5 | |
Detector Collimation (mm) | 2.5 | |
Slice Thickness (mm) | 3 or 5 | |
Data Reconstruction Interval (mm) | 3 or 5 | |
Table Speed (mm per rotation) / Pitch | 12.5/5 | |
Oral Contrast | N/A | |
IV Contrast Volume and Type | 120 cc of Omnipaque 350 | |
Injection Rate | 2 cc | |
Scan Delay (sec) | 40-50 sec | |
3D Technique Used | N/A | |
Comment: A second reconstruction of the data can be done with a bone algorithm (B60) for better bone detail. You can film the muscle/soft tissue with the reconstructions from the standard acquisition (B30-soft tissue) of the bone with the bone algorithm. |