Siemens Definition 64 Slice MDCT Protocols: Head : 1CTA_CAROTIDS_BOLUS_TRACKING (Adult)

Source: JHU Protocol
Protocol Range Series Description kV Quality Ref. mAs (Eff.) mAs CAREDose/CAREDose 4D CAREDoseType CTDIvol RotTime Pitch
1CTA_CAROTIDS_BOLUS_TRACKING (Adult)Topogram Topogram 0.6 T20s 120 0 327 off CAREDose 0 0.500 1.0000000
 PreMonitoring  120 120 60 off CAREDose4D 8.66065 0.330 0.6000000
   PreMonitoring 9.6 B20f        
 I.V. Bolus         
 Monitoring  120 120 60 off CAREDose4D 259.82 0.330 0.6000000
   Monitoring 9.6 B20f        
 CTA_NECK  120 120 300 off CAREDose4D 21.6176 0.330 0.70
   CTA_NECK 0.75 B20f -thins        
   CTA_NECK 3.0 B30f        
   CTA_NECK 3.0 B70f -Bone        
Coll. Slice Recon increment Images Kernel Window API Comment1
 0.60 0.60  1 T20s Topogram Body None 
 0.60      None 
  9.60  1 B20f Larynx  
        
 0.60      None 
  9.60  30 B20f Larynx  
 0.60      None 
  0.75 0.5 501 B20f Larynx  
  3.00 3.0 84 B30f Larynx  
  3.00 3.0 84 B70f NEURO BONE