| Specific anatomic region | Cardiac |
| Application | Calcium scoring sequence |
| Author | Elliot K Fishman, MD |
| Reference Source | JHU Protocol |
| Scanner Used | 64 Slice |
| KV/Effective mAs/ Rotation time (sec) | 120/ 200 / .36 |
| Detector Collimation (mm) | .6 |
| Slice thickness (mm) | 3mm |
| Pitch | 18mm |
| Kernel | B25 smooth + + |
| Reconstruction interval | 3mm |
| Image order (acquisition) | Cr-ca |
| Oral contrast | N/A |
| IV contrast volume and type | N/A |
| Injection rate | N/A |
| Scan delay (sec) | N/A |
| 3D technique used | N/A |
| Comment: Calcium scoring is then done with scores using Agaston, equivalant mass, and volume (mm3) provided. | |