| Specific anatomic region | Coronary arteries |
| Application | Bypass graft patency |
| Author | Elliot K Fishman, MD |
| Reference Source | JHU Protocol |
| Scanner Used | 64 Slice |
| KV/Effective mAs/ Rotation time (sec) | 120/ 800 / .33 |
| Detector Collimation (mm) | .6 mm |
| Slice thickness (mm) | .75 mm |
| Pitch | .2 |
| Kernel | B20 |
| Reconstruction interval | .4 mm |
| Image order (acquisition) | cr-ca |
| Oral contrast | N/A |
| IV contrast volume and type | 80-100 cc Visipaque / 40cc saline |
| Injection rate | 4cc/sec |
| Scan delay (sec) | test bolus |
| 3D technique used | VRT/MIP/MPR |
| Comment: When scanning grafts like LIMA grafts you must scan from the lung apex downward to get origin of the graft in most cases. | |