Specific anatomic region | Internal auditory canal (IAC) |
Application | R/O mass |
Author | Elliot K Fishman, MD |
Reference Source | JHU Protocol |
Scanner Used | 64 Slice |
KV/Effective mAs/ Rotation time (sec) | 120/ 200 / 1.0 |
Detector Collimation (mm) | 0.3 |
Slice thickness (mm) | .4 |
Pitch | 1 |
Kernel | 70 UHR |
Reconstruction interval | .2 mm |
Image order (acquisition) | Cr-ca |
Oral contrast | N/A |
IV contrast volume and type | N/A |
Injection rate | N/A |
Scan delay (sec) | N/A |
3D technique used | VRT & MPR |
Comment:
- The best in class for base of skull
|