64 Slice MDCT Protocols: Head and Neck |
|
|
Specific anatomic region |
Neck |
|
Application |
R/O mass or nodes or abscess |
|
Author |
Elliot K Fishman, MD |
|
Reference Source |
JHU Protocol |
|
Scanner Used |
Siemens Sensation 64 |
|
KV/Effective mAs/ Rotation time (sec) |
120/ 200 / 1.0 |
|
Detector Collimation (mm) |
0.6 |
|
Slice thickness (mm) |
3mm and .75mm (for 3D) |
|
Pitch |
0.9 |
|
Kernel |
B31f
medium smotth + : Larynx window B70s very sharp : Head Bone window |
|
Reconstruction interval |
3mm x 3mm ST and Bone & .5mm (3D) |
|
Image order (acquisition) |
Cr-ca |
|
Oral contrast |
N/A |
|
IV contrast volume and type |
100-120 cc Omnipaque 350 |
|
Injection rate |
3.0-2.5 cc/sec |
|
Scan delay (sec) |
Depends on patient, but in 30 sec range |
|
3D technique used |
N/A |
|