64 Slice MDCT Protocols: Chest

Specific anatomic region

Chest

Application

PE study

Author

Elliot K Fishman, MD

Reference Source

JHU Protocol

Scanner Used

Siemens Sensation 64

KV/Effective mAs/ Rotation time (sec)

100/ 200 / 0.33

Detector Collimation (mm)

0.6

Slice thickness (mm)

0.75mm

Pitch

0.75mm

Kernel

B20f smooth PE window

Reconstruction interval

0.5mm

Image order (acquisition)

Cr-ca

Oral contrast

N/A

IV contrast volume and type

100-120 cc Visipaque 320

Injection rate

3-3.5cc/sec

Scan delay (sec)

25-30 sec or test bolus

3D technique used

N/A

Comment: MPR or VRT may be helpful. Use InSpace and give it a try. I especially like the coronal MPR with a 3mm width.