64 Slice MDCT Protocols: Chest

Specific anatomic region

Chest

Application

Hi Res Chest (ILD)

Author

Elliot K Fishman, MD

Reference Source

JHU Protocol

Scanner Used

Siemens Sensation 64

KV/Effective mAs/ Rotation time (sec)

120/ 125 / 0.5

Detector Collimation (mm)

0.6

Slice thickness (mm)

1mm

Pitch

0.9

Kernel

B80f ultra sharp lung window

Reconstruction interval

10mm

Image order (acquisition)

Cr-ca

Oral contrast

N/A

IV contrast volume and type

If needed 120cc Omnipaque 350

Injection rate

2-2.5cc/sec

Scan delay (sec)

60 sec

3D technique used

VRT & MIP

Comment: When 3D is done reconstruction is .75mm at .5 mm intervals.