64 Slice MDCT Protocols: Chest

Specific Anatomic Region

Chest

Application

R/O Occupational Disease (asbestosis)

Author

Elliot K. Fishman, MD

Reference Source

JHU Protocol

Scanner Used

Siemens Sensation 64

KV/Effective mAs/Rotation time (sec)

120/150/0.5

Detector Collimation (mm)

0.6

Slice Thickness (mm)

3 and .75

Pitch

0.9

Kernel

30 (soft tissue) and 80 (lung)

Reconstruction Interval

2 and .5

Image Order (acquisition)

Cr-ca

Oral Contrast

N/A

IV contrast volume and type

None

Injection Rate

N/A

Scan delay (sec)

end inspiration

3D technique used

VRT and MIP as needed

Comment: The use of 3D with VRT and MIP is useful in my experience for looking at the lung parenchyma.