64 Slice MDCT Protocols: Abdomen |
|
|
Specific anatomic region |
Kidney |
|
Application |
R/o Kidney mass |
|
Author |
Elliot K Fishman, MD |
|
Reference Source |
JHU Protocol |
|
Scanner Used |
Siemens Sensation 64 |
|
KV/Effective mAs/ Rotation time (sec) |
120 / 250/ 0.33 |
|
Detector Collimation (mm) |
0.6 |
|
Slice thickness (mm) |
0.75mm (thin) 3.0mm (thick) |
|
Pitch |
0.75 |
|
Kernel |
B20f
smooth abdomen window (thin) B30f medium smooth abdomen (thick) |
|
Reconstruction interval |
0.5mm
(thin) 3.0mm (thick) |
|
Image order (acquisition) |
Cr-ca |
|
Oral contrast |
water (1000 cc) |
|
IV contrast volume and type |
120 cc Omnipaque 350 |
|
Injection rate |
3.0-3.5 cc/sec |
|
Scan delay (sec) |
Non:
4 sec Arterial: 30 sec Delay: 5 min |
|
3D technique used |
VRT & MIP |
|
Comment:
Also have protocol using 0.5 sec rotation time. |
|