64 Slice MDCT Protocols: Abdomen

Specific anatomic region

Abdomen/pelvis

Application

Trauma

Author

Elliot K Fishman, MD

Reference Source

JHU Protocol

Scanner Used

Siemens Sensation 64

KV/Effective mAs/ Rotation time (sec)

120 / 250/ 0.33

Detector Collimation (mm)

0.6

Slice thickness (mm)

3-5mm

Pitch

0.7

Kernel

30

Reconstruction interval

3-5mm

Image order (acquisition)

Cr-ca

Oral contrast

1000 cc water or Omnipaque-350 in water (100cc of Omnipaque-350 in one gallon of water)

IV contrast volume and type

100-120 cc Omnipaque 350

Injection rate

3cc/sec

Scan delay (sec)

30-60 sec & 5 minutes

3D technique used

as needed

Comment:

  • Scan delay may be shortened to look for active bleeding.
  • Delayed scans at 5 minutes are needed to exclude contrast extravasation from the kidney or bladder.