Radiology. 2015 Sep;276(3):706-14. doi: 10.1148/radiol.15140427. Epub 2015 Mar 17.
Geyer LL1, Glenn GR1, De Cecco CN1, Van Horn M1, Canstein C1, Silverman JR1, Krazinski AW1, Kemper JM1, Bucher A1, Ebersberger U1, Costello P1, Bamberg F1, Schoepf UJ1.
Purpose
To use suitable objective methods of analysis to assess the influence of the combination of an integrated-circuit computed tomographic (CT) detector and iterative reconstruction (IR) algorithms on the visualization of small (≤3-mm) coronary artery stents.
Materials and Methods
By using a moving heart phantom, 18 data sets obtained from three coronary artery stents with small diameters were investigated. A second-generation dual-source CT system equipped with an integrated-circuit detector was used. Images were reconstructed with filtered back-projection (FBP) and IR at a section thickness of 0.75 mm (FBP75 and IR75, respectively) and IR at a section thickness of 0.50 mm (IR50). Multirow intensity profiles in Hounsfield units were modeled by using a sum-of-Gaussians fit to analyze in-plane image characteristics. Out-of-plane image characteristics were analyzed with z upslope of multicolumn intensity profiles in Hounsfield units. Statistical analysis was conducted with one-way analysis of variance and the Student t test.
Results
Independent of stent diameter and heart rate, IR75 resulted in significantly increased xy sharpness, signal-to-noise ratio, and contrast-to-noise ratio, as well as decreased blurring and noise compared with FBP75 (eg, 2.25-mm stent, 0 beats per minute; xy sharpness, 278.2 vs 252.3; signal-to-noise ratio, 46.6 vs 33.5; contrast-to-noise ratio, 26.0 vs 16.8; blurring, 1.4 vs 1.5; noise, 15.4 vs 21.2; all P < .001). In the z direction, the upslopes were substantially higher in the IR50 reconstructions (2.25-mm stent: IR50, 94.0; IR75, 53.1; and FBP75, 48.1; P < .001).
Conclusion
The implementation of an integrated-circuit CT detector provides substantially sharper out-of-plane resolution of coronary artery stents at 0.5-mm section thickness, while the use of iterative image reconstruction mostly improves in-plane stent visualization. (©) RSNA, 2015