Protocols- Head and Neck

Circle of Willis- R/O Aneurysm


Specific Anatomic Region Circle of Willis
Application R/O aneurysm
Author Katz DA et al.
Reference Source Radiology 1995; 195:445-449
Scanner Used Siemens Somatom Plus-S
Injection Rate 3ml/sec
Contrast Volume and Type 90ml of Omnipaque-300
Area Scanned Entire brain (if possible)
Scan Delay Test-dose timing (usually 15-25sec)
Length of Spiral (time) 40sec
Slice Thickness 1mm
Table Speed/Pitch 1.3mm/sec with a pitch of 1.3
Reconstruction Interval 0.5mm
3D Technique Used None used
Comment: Spiral compared favorably with MRI. We like to use 3D imaging with volume rendering when looking at the aneurysm and its vascular anatomy. By using fly-around virtual techniques the neck of the aneurysm is best seen.