Protocols- Head and Neck

Carotid Artery Stenosis 2


Specific Anatomic Region Carotid Arteries
Application Carotid artery stenosis
Author Dillon et al.
Reference Source Radiology 1993; 189:211-219
Scanner Used Tomoscan SR-7000
Injection Rate 1.8ml/sec
Contrast Volume and Type 120ml of Ioxitalamate (Telebrix)-350
Area Scanned Base of skull till C5-6
Scan Delay 3.5sec
Length of Spiral (time) 40sec
Slice Thickness 3mm
Table Speed/Pitch 3mm/sec with a pitch of 1
Reconstruction Interval 1mm
3D Technique Used Shaded-surface rendering technique
Comment: Patients had their heads tilted back and the gantry was angled so that it was perpendicular to the presumed course of the carotid arteries. The published results with shaded-surface rendering have been poor and this 3D technique should be avoided. MIP or volume rendering would be preferred.