Protocols- Abdominal

Renal Arteries- Renal Artery Stenosis Evaluation


Specific Anatomic Region Renal arteries
Application Renal artery stenosis evaluation
Author Galanski M et al.
Reference Source Radiology 1993; 189:185-192
Scanner Used Siemens Somatom Plus or Plus-S
Injection Rate 3-5ml/sec (choice 3.5ml/sec)
Contrast Volume and Type 100-150ml of Iopromide-300
Area Scanned Entire length of the kidneys
Scan Delay Test-dose timing (usually 20-30sec)
Length of Spiral (time) 32sec
Slice Thickness 2mm
Table Speed/Pitch 2mm/sec with a pitch of 1
Reconstruction Interval 1-2mm
3D Technique Used Gradient shaded-surface rendering (with threshold of 130-220 HU) and MIP
Comment: Test dosing to optimize timing was used in this study. This is similar to work by Rubin et al. but they inject at 5cc/sec. A key in these cases is to look for accessory renal arteries which increase the length of coverage needed.