| Specific Anatomic Region | Liver |
| Application | Liver transplant evaluation |
| Author | Smith P et al |
| Reference Source | AJR (in press) |
| Scanner Used | Siemens Somatom Plus-4 |
| Injection Rate | 3cc/sec |
| Contrast Volume and Type | 150 |
| Area Scanned | From just above dome of liver to inferior margin of liver |
| Scan Delay | 30sec / 60-70sec |
| Length of Spiral (time) | 30-35sec dual spiral |
| Slice Thickness | 3mm |
| Table Speed/Pitch | 5-6mm/sec |
| Reconstruction Interval | Every 1mm for arterial phase and every 3mm for venous phase |
| 3D Technique Used | Volume rendering |
| Comment: Dual phase scanning provides evaluation of the liver parenchyma, arterial supply and venous anatomy. | |